※一部利用できない機能があります
2点相関測定用アダプタ付き光ファイバLDAプローブ
- フォーマット:
- 論文
- 責任表示:
- 小保方, 富夫 ; Bopp, Stefan ; Tropea, Cam
- 言語:
- 日本語
- 出版情報:
- 日本機械学会, 1989-05-25
- 著者名:
- 掲載情報:
- 日本機械学會論文集. B編
- ISSN:
- 0387-5016
- 巻:
- 55
- 通号:
- 513
- 開始ページ:
- 1490
- 終了ページ:
- 1493
- バージョン:
- VoR
- 概要:
- application/pdf<br />Journal Article<br />A simple adapter for a two-dimensional, backscatter, fiber-optic laser Droppler anemometer probe consisting of two glass plates has been developed to measure the spatial correlation of velocity fluctuations. A conventional two-color, four-beam LDA probe has been modified by reversing the placement of two of the fibers so that both the green and the blue control volumes measure the same velocity component. … An adapter for beam separation is ten mounted behind the front lens to separate the two control volumes. In the present work, the measuring point separation was varied up to 3 mm in the lateral and vertical directions. By using only one glass plate 10 mm thick, it was also possible to achieve in-line (Axial) displacements of the control volumes of up to 3.5 mm. spatial correlations were performed in an intermittent jet flow in which the convective flow velocity could be deduced from the offset of the cross-correlation function. 続きを見る